Characterization and Fatigue of Spinal Intervertebral Body Fusion Devices (ASTM F2077)

Intervertebral Body Fusion devices are used to promote arthrodesis of a spinal motion segment, following disc degeneration. The implant acts as a support of the anterior column whilst the two vertebral discs grow together and fuse the spine. Intervertebral Body Fusion is an alternative technique to that of Intervertebral Disc replacement.

ASTM F2077 – 03 “Test Methods for Intervertebral Body Fusion Devices” specifies a number of different static and dynamic tests to provide a mechanical comparison between different fusion devices. The tests include:

  • Axial-Compression
  • Compression-Shear
  • Torsion Testing

An axial-torsion system, such as the 8874 servohydraulic or the ElectroPuls™ E10000 Linear-Torsion All-Electric Dynamic Test Instrument, allows manufacturers and researchers to conduct both the static and fatigue testing on a range of implant designs. These systems are combined with a temperature controlled bath, which provides a stable environment for simulation of in vivo conditions.

With the use of specialized test fixtures, the combined axial-torsional actuator in the crosshead of either the 8874 or the E10000 allows for conductin characterization in axial compression, compression-shear, or compression-torsion tests modes. Our testing systems allow for mounting of the Biaxial Dynacell™ load cell on the end of the moving axial-torsional actuator and can automatically compensate for errors caused by inertial loading.

Review alternative test for intervertebral body fusion devices is specified in ASTM F2267.

We recommend that you review the standard to fully understand its requirements.

Are you interested in other types of test on spinal implants and spinal devices? Details can be found here.


About This Solution

Related Standards: ASTM F2077 | ASTM F2267
Specimen Type: Implant
Materials: Biomedical | Biomaterials
Type of Test: Compression (What is a Compression Test?) | High-Cycle Fatigue | Multi-Axis | Torsion | Shear
Business Sector: Academic | Biomedical/Medical/Healthcare | Contract Testing Services | Government/Defense
Test Fixture for Spinal Interbody Fusion Devices to ASTM standards
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
Interested in this Solution?
Contact Us

Related Literature:

ElectroPuls™ Electrodynamic Test Instruments   ElectroPuls™ Electrodynamic Test Instruments
Fast Track 8870 Series Servohydraulic Fatigue Test Systems   Fast Track 8870 Series Servohydraulic Fatigue Test Systems
Solutions for Biomedical Testing   Solutions for Biomedical Testing

Our Labs

Instron application labs are used for both research and development, as well as testing customer samples to assist in finding the right equipment.
Contact our Application Specialists to determine the solution that's best for you!